Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 5725A: (90 day) CAL VER ACV IEEE /5790 DATE: 26-Aug-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.9 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 10 NUMBER OF LINES: 114 CONFIGURATION: Fluke 5790A ============================================================================= # # Source: # Fluke 5700A Service Manual, Section 3, Rev 2, May 1991 # # Compatibility: # MET/CAL 5.0 or later # # Subprocedures: # Sub Wait for Operation Complete (OPC) IEEE # # Required Files: # 5790CON1.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5790A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5790A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R Q N F W # Set "Wait for OPC" timeout to 30000ms. 1.002 MATH M[88] = 30000 1.003 JMPT 1.016 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this verification. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 OPBR Is the Fluke 5700A to be calibrated a configured system 1.006 OPBR instrument connected to IEEE-488 port 0 on this PC? 1.007 JMPT 1.012 1.008 DISP Connect an IEEE-488 cable from the 5700A to 488 Port 1. 1.009 DISP Turn on the 5700A and allow 30 minutes warmup time. 1.010 DISP Allow the Fluke 5700A to stablize in an environment 1.010 DISP with ambient temperature of 21 to 25 C and relative 1.010 DISP humidity of less than 70%. 1.011 JMP 1.013 1.012 DISP Enter the address of the 5700A (MET/CAL default = 15) 1.012 DISP when prompted for the UUT address. 1.013 PIC 5790CON1 1.014 RSLT =ACV CALIBRATION 1.015 RSLT = 1.016 IEEE *CLS;*RST 1.017 CALL Sub Wait for Operation Complete (OPC) IEEE 1.018 JMPL END M[89] <= 0 1.019 IEEE BTYPE VB5725 1.020 ACMS 300V M 1.021 HEAD ACV CALIBRATION {1100V RANGE (5725A)}: 300V @ 20kHz 1.022 IEEE [T30000]OUT 300V,20kHz;OPER 1.023 CALL Sub Wait for Operation Complete (OPC) IEEE 1.024 JMPL END M[89] <= 0 1.025 5790 1100 300.000V 155P% 20kH I1 2W 2.001 ACMS 600V M 2.002 HEAD ACV CALIBRATION 1100V RANGE (5725A): 600V @ 50kHz 2.003 IEEE [T30000]OUT 600V,50kHz;OPER 2.004 CALL Sub Wait for Operation Complete (OPC) IEEE 2.005 JMPL END M[89] <= 0 2.006 5790 1100 600.000V 378P% 50kH I1 2W 3.001 HEAD ACV CALIBRATION 1100V RANGE (5725A): 600V @ 100kHz 3.002 IEEE [T30000]OUT 600V,100kHz;OPER 3.003 CALL Sub Wait for Operation Complete (OPC) IEEE 3.004 JMPL END M[89] <= 0 3.005 5790 1100 600.000V 0.1375% 100kH I1 2W 4.001 ACMS 1000V M 4.002 HEAD ACV CALIBRATION 1100V RANGE (5725A): 1000V @ 40Hz 4.003 IEEE [T30000]OUT 1000V,40Hz;OPER 4.004 CALL Sub Wait for Operation Complete (OPC) IEEE 4.005 JMPL END M[89] <= 0 4.006 5790 1100 1000.000V 84P% 40H I1 2W 5.001 HEAD ACV CALIBRATION 1100V RANGE (5725A): 1000V @ 50Hz 5.002 IEEE [T30000]OUT 1000V,50Hz;OPER 5.003 CALL Sub Wait for Operation Complete (OPC) IEEE 5.004 JMPL END M[89] <= 0 5.005 5790 1100 1000.000V 84P% 50H I1 2W 6.001 HEAD ACV CALIBRATION 1100V RANGE (5725A): 1000V @ 300Hz 6.002 IEEE [T30000]OUT 1000V,300Hz;OPER 6.003 CALL Sub Wait for Operation Complete (OPC) IEEE 6.004 JMPL END M[89] <= 0 6.005 5790 1100 1000.000V 84P% 300H I1 2W 7.001 HEAD ACV CALIBRATION 1100V RANGE (5725A): 1000V @ 1kHz 7.002 IEEE [T30000]OUT 1000V,1kHz;OPER 7.003 CALL Sub Wait for Operation Complete (OPC) IEEE 7.004 JMPL END M[89] <= 0 7.005 5790 1100 1000.000V 84P% 1kH I1 2W 8.001 HEAD ACV CALIBRATION 1100V RANGE (5725A): 1000V @ 20kHz 8.002 IEEE [T30000]OUT 1000V,20kHz;OPER 8.003 CALL Sub Wait for Operation Complete (OPC) IEEE 8.004 JMPL END M[89] <= 0 8.005 5790 1100 1000.000V 131P% 20kH I1 2W 9.001 HEAD ACV CALIBRATION 1100V RANGE (5725A): 1000V @ 30kHz 9.002 IEEE [T30000]OUT 1000V,30kHz;OPER 9.003 CALL Sub Wait for Operation Complete (OPC) IEEE 9.004 JMPL END M[89] <= 0 9.005 5790 1100 1000.000V 371P% 30kH I1 2W 10.001 LABEL END 10.002 IEEE [T0]*RST 10.003 5790 * S 10.004 END